<WRAP left round info 50%> | <WRAP left round info 50%> | ||
- | Notice: It is very easy to warp dichroic mirrors; only lightly tighten the retaining clip of the D-CUBE. | + | Notice: It is easy to warp dichroic mirrors |
</ | </ | ||
\\ \\ \\ \\ \\ | \\ \\ \\ \\ \\ | ||
Line 186: | Line 186: | ||
===== Beam position ===== | ===== Beam position ===== | ||
- | We are concerned with getting both the beam position and angle correct. Fortunately, | + | We are concerned with getting both the beam position and angle correct, along with getting both objectives " |
< | < | ||
^ Adjustment | ^ Adjustment | ||
- | | Tilt of dichroic mirror | + | | Tilt of dichroic mirror |
- | | Tilt of camera mirror | + | | Tilt of camera mirror |
- | | Position of upper Z stage | No | No | Angle varies if scanners are not connected to dichroic tubes | + | | Z position of SPIM head | No | No | |
| Objective bushing (focus) | | Objective bushing (focus) | ||
| Linear objective adjustment | | Linear objective adjustment | ||
- | | < | + | | < |
- | | < | + | | < |
</ | </ | ||
Line 248: | Line 248: | ||
**Outline**\\ | **Outline**\\ | ||
+ | * Phase 0 (coarse alignment by eye as described above) | ||
* Phase 1 (iterative) | * Phase 1 (iterative) | ||
- | * 1. Co-align SPIM objectives to focus and center beams. | + | * 1.1 Co-align SPIM objectives to focus and center beams. |
- | * 2. Tilt dichroic mirrors for horizontal | + | * 1.2 Tilt dichroic mirrors for __horizontal__ |
- | * 3. Tilt imaging mirrors to center epi spots. | + | * 1.3 Tilt imaging mirrors to center epi spots. |
+ | * 1.4 Adjust lateral position of objective. | ||
+ | * 1.5 Go back to 1.1 and repeat until no adjustments needed | ||
* Phase 2 (may have to be repeated for initial alignment, and again when scanners or cameras are changed or disturbed) | * Phase 2 (may have to be repeated for initial alignment, and again when scanners or cameras are changed or disturbed) | ||
- | * 4. Focus collimators to center beam waists in FOV. | + | * 2.1 Focus collimators to center beam waists in FOV. |
- | * 5. Rotate scanners for __uniform__ focus of the sheets. | + | * 2.2 Rotate scanners for __uniform__ focus across |
- | * 6. Rotate cameras for vertical sheets from epi view. | + | * 2.3 Rotate cameras for vertical sheets from epi view. |
- | * Phase 3 (performed just once for initial alignment and again for objective | + | * Go back to 1.1 |
- | * 7. Establish the coverslip location. | + | * Phase 3 (performed just once for initial alignment and again for objective |
- | * 8. Align the inverted scope with the SPIM objectives. | + | * 3.1 Establish the coverslip location. |
- | * 9. Cross-calibrate the piezo and light sheet movement. | + | * 3.2 Align the inverted scope with the SPIM objectives. |
+ | * 3.3 Cross-calibrate the piezo and light sheet movement. | ||
+ | * Phase 4 (performed for any new sample) | ||
+ | * 4.1 Check that light sheet and detection plane are coincident; adjust offset in plugin if needed. | ||
==== Phase 1 ==== | ==== Phase 1 ==== |